The nuclear medicine ventilation fume hood is equipped with an independent air supply and exhaust system to meet the requirements of dispensing, elution and radiolabeling.
Email: admin@cnrayshield.com
Specifications: External : 1200(w)×750(d) ×2100(h)mm
Inner size : 950(w)x580(d) x 670(h)mm
1.Lead equivalent: 40mmPb. ( Option 2: I-131 40 mm Pb on the front&bottom,30 mm on the sides, 20 mm on the back&top )
2.The front of the cabinet is equipped with a lead glass observation window
(230×600mm), which has the same protection capacity as the cabinet.
3.There are operation holes on the front, equipped with gloves.
4.There are delivery doors on both sides of the cabinet
5.The built-in calibrator protection well is designed according to the
scintidose calibrator CRC—55TR(D200x H420mm) , lead equivalent 40mmPb.
6.Built-in molybdenum-technetium generator automatic lifting system
(lead equivalent 50mmpb).
7.Equipped with lighting , sterilization , ventilation , high-efficiency filter ,
power supply devices, power supply 50HZ/20A/220V with safety protection,
centralized control and adjustment panel.
8.Inside and outside material with SS304 steel, inside mirror surface for easy views
9.The radius treatment for the inner tank ensures the cleanliness of the inside cabinet.
10.Configure vacuum negative pressure gauge. Vacuum value 0~250Pa
11.The bottom of the cabinet is equipped with universal casters.
PET(FDG) | HS-07-S-TFG | 40mmPb |
I-131
|
HS-08-S-TFG |
40 mm Pb on the front&bottom,30 mm on the sides, 20 mm on the back, top and bottom |
SPECT(Tc99) | HS-06-S-TFG | 10mmPb |
Please Leave your message you want to know! We will respond to your inquiry within 24 hours!